for further technical information, please contact advanced thermal solutions, inc. at 1-781-769-2800 or www.qats.com 89-27 access road, norwood, ma 02062 usa | t: 781.769. 2800 f: 781.769.9979 | www.qats.com r2_0215 cold plate characterization system application domain ? evaluating and troubleshooting different cold plate models ? simulating different applications and conditions ? optimizing fow and thermal performance ? testing alternative liquids IFLOW-200 tm measure the thermal & hydraulic performance of cold plates overall dimensions (d x w x h) 365 mm x 425 mm x 120 mm maximum volumetric flow rate 4.5 liter/min flow temperature range 0 to 70 o c surface temperature range 0 to 200 o c differential pressure range +/- 103,000 pa(15 psi) pressure accuracy +/- 1% of full scale temperature accuracy +/- 1 o c flow rate accuracy +/-1% of full scale software cooling view tm ? IFLOW-200? measures a cold plates inlet/outlet fuid temperature, surface temperature, coolant volumetric fow rate and pressure drop by providing different fow rates to the cold plate. ? IFLOW-200? measures coolant temperature from 0-70 o c with accuracy of +/-1 o c, using k-type thermocouples. distilled water is used as the reference coolant. ? IFLOW-200? measures cold plate differential pressure up to 103,000 pa (15 psi) with pressure measurement of +/-1% of full scale. ? IFLOW-200? features a separate control box and hydraulics units with usb connections. ? IFLOW-200? includes user-friendly coolingview? application software that automates the testing of a cold plate by setting the required parameters. ? IFLOW-200? software, coolingview tm , calculates thermal resistance and pressure drop of the cold plate as a function of fow rate for selected liquids other than distilled water. fluid level indicator inlet/outlet of the cold plate sensor ports cooling system for the internal heat exchanger features & benefits the IFLOW-200 tm system measures the thermal and hydraulic characteristics of a cold plate. it replaces the complex process of varying the volumetric fow rate of the coolant, and measuring the pressure drop and temperature of the cold plate. IFLOW-200 tm users simply set up the starting fow, ending fow, number of test points, dwell time, power applied to the cold plate and other parameters in the coolingview tm software and run the application. features: ? free lifetime tech support
|